34
Mecmesin
MultiTest
-dV
Operating Manual
AFG Control Test
With an additional cable (351-092), a Mecmesin AFG digital force gauge can be used to
set load limits to control crosshead movement.
Load setpoint, action (reverse/stop) are all set on the gauge under the ‘STAND’ menu.
Here you can select the action when the limit is reached; ‘REVERSE’ or ‘STOP’ and the
type of control limit ‘BREAK’ or ‘LIMIT’.
Please note for cyclic tests the front panel force control must be used. AFG cycling is not
compatible with MultiTest-
dV
test stands
Example Test Setup
1.
On the test stand’s front panel under ‘Edit Test’ select the test type ‘AFG Control’
and sub-type ‘AFG Control’,
2.
Configure the speed and direction settings located within the ‘Edit Test’ menu,
3.
On the AFG gauge hold
4.
On page one of the AFG menus select ‘STAND’ using
the button,
5.
Next select the action when the limit is hit, either ‘REVERSE’ or ‘STOP’. For
reverse you need to select the test direction either ‘UP’ or ‘DOWN’,
6.
Once the action has been selected configure the limit control. This limit can be
either a ‘BREAK’ condition or force/torque ‘LIMIT’. For ‘BREAK’ set the break
threshold, for ‘LIMIT’ select the limit force/torque,
Force Control Test – Force Limit, Force Cycle and Move to Break
Force control tests can be used to set load limits or a break condition to control crosshead
movement.
Within the three sub-tests (Force Limit, Force Cycle and Move to Break) the following
settings are available, please note some of the settings are specific to the test type:
Setting
Options
Up Speed
0-1200mm/min
Down Speed
0-1200mm/min