Press the
RESET
RESET
key to clear both maximum registers and prepare for detecting the next maximum readings.
Data Output
The BFG has RS232, Mitutoyo and analogue output signals. It is possible to transmit the displayed reading to
peripheral devices (e.g. PC, printer) by pressing and releasing the
TXD
TXD
key.
Displayed readings can also be requested individually from a PC via the RS232 interface by sending a “
??
” (ascii
D63 [3fh]) character.
For sending a continuous data stream to a PC, press and hold the
TXD
TXD
key for 2 seconds then release. A ‘1’ will
now appear in the display to indicate that data can now be sent. To stop sending data, simply press and release
the
TXD
TXD
key, at which point a ‘1’ will disappear from the display.
Please note that the continuous data stream only starts when approximately 2% of the rated capacity of the gauge
is reached.
BFG uses 9600 Baud 8 data bits, 1 start bit 1 stop bit and no parity.
BFG uses 9600 Baud 8 data bits, 1 start bit 1 stop bit and no parity.
When the continuous data stream option is selected the gauge defaults to a BAUD rate of 57,600.
When the continuous data stream option is selected the gauge defaults to a BAUD rate of 57,600.
A full range of data cables are available to connect your gauge to peripheral devices – see page 12.
A full range of data cables are available to connect your gauge to peripheral devices – see page 12.
Remote key press from PC
Hold down the Ctrl key on the keyboard and pressing the following keys or sending their equivalent ASCII code
as shown in brackets:
a
a
to simulate pressing the
TXD
TXD
key* (ascii DØ1 [Ø1h])
b
b
to simulate pressing the
UNITS
UNITS
key (ascii DØ2 [Ø2h])
c
c
to
simulate pressing the
MAX
MAX
key (ascii DØ3 [Ø3h])
d
d
to simulate pressing the
RESET
RESET
key (ascii DØ4 [Ø4h])
e
e
to simulate pressing the
ZERO
ZERO
key (ascii DØ5 [Ø5h])
Note
Note
that the continuous transmission mode cannot be entered via this method
Factory Defaults
The BFG is delivered from the factory with the following settings as default:
Display: “Normal” mode
Auto-off: Not active
Transmit minus sign: Not active
Optional Settings through Dual Function keys
Auto-off
To conserve battery power, it is possible to activate an Auto-off function so that the gauge powers down after 5
13