PSE
.
17
.
CAL Submenu:
CAL
SUBMENU1
SUBMENU2
OPTION
REMARK
CAL.ON
CAL.OFF
Seal switch is on or off
ZERO
Only perform zero point calibration, then go to
CAL.END
to end
LINE
CAL.P0
Linear calibration point0: perform zero point
calibration, this point can not be omitted.
CAL.P1
Linear calibration point1: complete the first weight
point calibration, this point can not be omitted and
standard weight must be over 10%FS.
END.Y
YES
End calibration?
YES
=go to
CAL.END
to end;
NO
=perform the next point calibration
NO
CAL.P2
Linear calibration point2: complete the second
weight point calibration, standard weight
must be
over 10%FS and must be larger than it in
CAL.P1
, this point can be omitted.
END.Y
YES
End calibration?
YES
=go to
CAL.END
to end;
NO
=perform the next point calibration
NO
CAL.P3
Linear calibration point3: perform third weight point
calibration, standard weight must be over 10%FS
and must be larger than it in
CAL.P2
, this point
can be omitted.
GEO
CODE
00-70
Selection of Geographical Position Code 00-70
GRAVT
9.76183
-9.99999
Input Gravity of User Location by keyboard
INPUT
Input or view calibration parameters value
CAL.END
Calibration end and restart
NOTE:
For more details, please refer to section
¡ 6.CALIBRATION¡