10. Err20
:
There is an error in calibration
11. Err30
:
ADC is over max. range
12. Err31
:
ADC is below min. range
13. Err40: Recall error
,
no content stored.
14. CAP.
:
Capacity
15. UOL.: Voltage
16. St.PLU: To set and store the indirect unit price
17. Addr.: PLU address
18. Unit: Weighing unit selecting
19. Rc.PLU: Recall indirect unit price
20. UnLoAd: To unload the weight
21 LoAd: To load the weight
22 InP.Ld: To input load weight
23 CAL.oN: Calibration enable switch is ON
24 CAL.oFF: Calibration enable switch is OFF
20.Troubleshooting and Battery charging
Troubleshooting
Troubleshooting
Possible causes
Solution
Power-on
problem
No display after press
the button
Not well load with the batteries or no
electricity or not well insert with adapter or
plug.
Check if the batteries have been load
with the wrong connection, if well
connection with batteries; Replace
new batteries; Check the power
supply with the adapter
there is something wrong with the electric
components of power on PCB,
crystal-oscillator or IC
Replace with new PCB
Key button not work
Replace with new key button
Display irregular
character after press
button
there is something wrong with the
crystal-oscillator on PCB, or MCU not well
insert
Replace the crystal-oscillator; well
insert the MCU.
Display 888888,not
back to 0
Something wrong with the keys, or faulty PCB
board
Check the key button; Replace the
PCB
Err
Err04
0 point exceed the 10% of the data when
calibrating; There are objects on the platform;
Load cell 0 point changes;
Remove the objects on the platform;
calibrate again.
Err05
0 point exceed the 10% of the data when
calibrating; There is one foot not stand on the
solid base; Load cell 0 point changes;
Lay the foot in the same level;
Calibrate again.