OPH-D/P/T/Z Series

User Manual

Summary of Contents for OPH-T3001

Page 1: ...OPH D P T Z Series User Manual ...

Page 2: ...rial onto the platform as that could damage the balance Do not spill liquids on the balance as that could damage the balance Material that has a static charge could influence the performance of the balance Electromagnetic interference can affect the reading of the balance Unplug the power before cleaning and use a piece of soft cloth do not use any aggressive cleaning agent Make sure no liquid ent...

Page 3: ... Preparation 6 Calibration 7 Using Scale 8 Basic Weighing 8 Tare 8 Changing Units 8 Counting Function 9 RS232 Communication Optional 9 More Info 10 Troubleshooting 10 Technical Support 10 OPH D P T Z Series User Manual Table of Contents 3 10 ...

Page 4: ...ccuracy grams OPH D202 200 0 01 OPH D302 300 0 01 OPH D502 500 0 01 OPH D602 600 0 01 OPH D2001 2000 0 1 OPH D3001 3000 0 1 OPH D5001 5000 0 1 OPH D6001 6000 0 1 OPH P103 100 0 001 OPH P203 200 0 001 OPH P303 300 0 001 OPH T202 200 0 01 OPH T302 300 0 01 OPH T502 500 0 01 OPH T602 600 0 01 OPH T2001 2000 0 1 OPH D P T Z Series User Manual Getting Started 4 10 ...

Page 5: ...Model Max Weight grams Accuracy grams OPH T3001 3000 0 1 OPH T5001 5000 0 1 OPH T6001 6000 0 1 OPH Z10K 10K 0 1 OPH Z20K 20K 0 1 OPH Z30K 30K 0 1 OPH D P T Z Series User Manual Getting Started 5 10 ...

Page 6: ...ale by using the bubble level indicator if available and adjusting the feet of the scale 5 Locate scale in the following condition 1 Temperature 12 5oC 27 5oC 2 Temperature Fluctuation 5oC h 3 Humidity 50 80 6 Plug in the scale to a surge protected power outlet 7 Press the ON OFF key in the front panel to turn the scale on and wait for it to finish initialization i e when the display shows a weigh...

Page 7: ...ion by testing different known weights 7 Re calibrate if measurements are found to be inaccurate Linear calibration 1 Verify the pan is empty 2 Press the CAL calibration key for about 5 seconds to enter into calibration mode When display shows CAL release the CAL key 3 Press and hold the MODE key for about 5 seconds and the blinking number indicates what weight should be placed on the pan 4 Load t...

Page 8: ... as possible 2 Read the value on the display when the reading is stable 3 Remember NOT to overload the scale since this may permanently damage the scale Tare 1 Make sure the scale is stabilized 2 Press theTARE key Changing Units Press the UNIT key to change between the following units CT caret OZ ounce lb pound g gram default OPH D P T Z Series User Manual Using Scale 1 8 10 ...

Page 9: ...he number of samples selected from step 4 on the weighing pan and press the MODE key 6 This will put the scale into the counting mode 7 Press the MODE key again to exit the counting mode back to the weighing mode RS232 Communication Optional RS232 port Parameters Data format 10 bit data frame bit1 bit2 bit 10 bit1 start bit bit 2 9 data bits bit 10 stop bit Baud rate 9600 bps All data in ASCII cod...

Page 10: ...libration procedure has been performed correctly It won t hurt to run through the calibration process again when in doubt Q The scale does not turn on A Make sure the power cord is plugged in and that there is power One easy way to test this is by connecting another appliance to the same outlet and see if it s operational Technical Support Please email support optimascale com if you have any furth...