9.
Troubleshooting
Troubleshooting
SYMPTOM
PROBABLE CAUSE
REMEDY
Err01
Weight reading exceeds the overload limit or the
weight value cannot be displayed in the current
unit of measure.
Reduce load on scale until weight value can
be displayed, or use an alternate unit of
measure.
Err03
Weight reading below minimum load limit.
Install platform on scale. Perform zero
calibration.
Err04
Weight exceeds Power On Zero limit (+15%), or
over ZERO key range (+5%).
Make sure scale platform is empty.
Power off scale and power on again.
Perform zero calibration.
Err05
Weight is below Power On Zero limit (-15%) or
below ZERO key range (-5%).
Install platform on scale.
Perform zero calibration.
Err30
Err31
Load cell wires to indicator are incorrectly
connected, shorted, or open; or ADC, load cell
are damaged.
Make sure wires are correctly connected.
Service required to replace load cell or ADC
chip.
Err10
EEPROM is damaged.
Service required to replace main EEPROM IC
chip.
Err11
Err12
Setup parameters are not set, not correctly set,
or settings have been lost.
Re-set parameters.
Perform calibration.
Err20
Calibration error. Input data or loaded weight is
too small, too large, unstable, or not linear.
Input correct data, load correct weight onto
platform, or service is required.
Err40
Data memory is not set, not correctly set, or set
data has been lost.
Reset data memory for the address you are
attempting to recall.
Will not power on
Power cord not plugged in or properly
connected. Power outlet not supplying electricity.
Battery discharged. Other failure.
Check power cord connections. Make sure
power cord is plugged into the power outlet.
Check power source. Replace batteries. Or
service required.
Unable to zero the
display or will not
zero when turned on
Load on scale exceeds allowable limits.
Load on scale is not stable.
Load cell damage.
Remove load on scale.
Wait for load to become stable.
Service required.
Incorrect counting
result when using
SPL to enter a piece
weight
Sampling quantity is too small.
Calculated piece weight is slightly different from
the real value.
Increase the sampling quantity.
Lo.bAt is shown
Battery is discharged
Charge battery
EHC
‐
C
www.measuretek.net
.