Symbol Definitions
Err01
- Weight signal is too large or over range
Err02
- No proper data can be displayed
Err03
- Weight signal is too small
Err04
- Zero point is over the setting range
Err05
- Zero point is below the setting range
Err10
- EEPROM cannot be accessed
Err11
- Parameters in EEPROM ≠ backup data
Err12
- Parameters in EEPROM are out of normal range
Err20
- There is a calibration error
Err30
- Input signal is over ADC’s max. range
Err31
- Input signal is below ADC’s min. range
Err40
- Recall memory data error (address is empty)
CAP.
- Capacity data
VoL.
- Voltage data
Add
- Accumulation data
PC.t
- Piece weight data
St.PC.t
- Set and store piece weight
Addr.
- Memory unit storage address
Unit
- Weighing unit
rc.PC.t
- Recall the stored piece weight, tare, and unit
Lo.PC.t
- Below minimum piece weight limit
SPL.PCS
- Sample piece data
Hi.PCS
- Upper limit piece data
Lo.PCS
- Lower limit piece data
Lo.SP.t
- Below minimum sample weight limit
UnLoAd
- Unload the weight
LoAd
- Load the weight
inP.Ld
- Input the load weight
dSP.ovr
- Data length exceeds the LED display digits
CAL.oN
- Calibration enable switch is ON
CAL.oFF
- Calibration enable switch is OFF
EHC
‐
C