CONVERTER INSTALLATION
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
for current technical data and instructions.
www.mccrometer.com
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA
Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078
Lit. # 30122-46 Rev. 1.2 / 5-31-18
5
2.3
Pulling Sensor Cable Through Electrical Conduit
It is very important to protect the end of the sensor cable when pulling it through a conduit. Water can accumulate
in low portions of conduit. Always use the factory supplied cable cover, or similar method, to seal the end of the
cable against water when pulling the cable through conduit. See Figure 4. This will ensure proper operation of
the meter.
Pulling The Sensor Cable:
1. Tie a rope or cable-snake securely around the middle of the cable cover.
2. Carefully pull the rope or snake until the sensor cable end clears the conduit.
3. Bring the cable end to the converter location. If necessary, secure the cable so that it does
not fall back through the conduit.
4. Remove the cable cover by pulling the rip wire. The cable cover will tear off (discard the cover).
CAUTION:
Do not cut the cable cover off. Doing so may damage the sensor cable and adversely effect the
calibration of the meter.
i
Sensor Cable
Cable Cover
Secure rope or snake to this
area of the cable cover
Rip Wire
Figure 4. Cable Cover