Page 20
-
BOTTLE FILLING PAUSE
(SÜV program )
Setting range: between 0-10080 min. (7 days).
This function enables a deferred sample extraction
(filling pause) related to the bottles before the next bottle
is filled.
The pause is entered in minutes.
Example
:
Sampler with 24-bottle distributor system.
Bottle fill time: 2 hours.
-
without
programmed bottle filling pause:
A bottle change is made every 2 hours, that means
after
24 hours
all the 12 bottles are filled.
-
with
programmed
bottle filling pause
of
24 hours
(1440
min.):
Bottle 1 is filled for 2 hours. Thereafter there is a bottle filling pause of 24
hours and only then the distributor changes to bottle No. 2.
Thus there is a delayed bottle filling of 24 hours between each of the individual
bottles. The whole program cycle in this example would be: 12 bottles x 26 h =
312 hours (2 hours fill time + 24 hours filling pause).
As a result of this setting, each bottle is filled with a delay of 26 hours per day.