FREELY PROGRAMMABLE INPUTS:
PROGR. INPUT
ADDITIONAL INPUT 1
ADDITIONAL INPUT 2
ADDITIONAL INPUT 3
There is
1
programmable input in the basic version.
With the option "I/O extension board" further
3
inputs are available.
Each input can be programmed individually according to the following list:
• NO FUNCTION
reset of setting
• PROGRAM START PULSE
if selected, program x can be started (
external start)
• PROGRAM STOP PULSE
if selected, program x will be stopped (terminate with ESC)
• PROGRAM RUN DURING PULSE
A program is executed as long as there is a continuous signal. If the signal
drops, the program will be stopped.
• BOTTLE CHANGE PULSE
Pulse signal:
<= 3sec
means “advance to next bottle”
>= 5 sec
means “move to bottle No. 1”
• SAMPLING PULSE
A pulse triggers the sample extraction
The pulse signal has to be > 50 ms!
Remark
:
This function is only possible if
no
program is running. In this case
the unit is controlled externally (e.g. via SPS).
• MANUAL SAMPLE
A manual sample is triggered.
There is no registration in the info memory (ideal for official samples or tests).
• ACKNOWLEDGE ERROR
Accumulated error messages can be acknowledged.