SETTINGS
DATE/TIME
Setting of date/time
DEVICE SETTINGS
• LANGUAGE
Setting of the language
• DISTRIBUTOR
Selection of distributor type
• MAX. SUCTION TIME
Setting of the maximum suction time
(0-600 sec.)
• PRE-PURGE
Pre-purge = purge of suction hose
PRIOR
of the sample extraction (0 - 99,99 sec.)
• POST-PURGE
Post-purge = active purging of the metering
vessel
AFTER
the sample extraction
(0 - 99,99 sec.)
• VAR CALIBRATION
Calibration of the system for flow-proportional
sampling
(only with VAR device option)
• AERATION TIME
Time until pinch valve is opened for drain off of
sample
• PUMP POWER
Adjustable from 70 % to 100%
• INTERNAL TEMPERATURE
-Via NTC
-Via PT1000
-Limit value (1 - 20 °C)
-Delay time (1 - 60 min.)
(Example: limit value 7°C, delay time 10 min. An
alarm message is sent if the limit value is
exceeded for 10 min.)
• ANALOG SIGNAL
Selection:
0-20 mA
4-20 mA
Calibration (adjustment with signal of plant)
• DISPLAY
- always switched on
- switch off after certain time (0-999 sec.)
- contrast
- max. brightness
- min. brightness
•DURATION OF PAUSE
Program can be interrupted for 10-120 min. for
example for cleaning .After expiry of the time
entered the program is automatically resumed.
• PROG. INPUTS
Programmable inputs
: input signal to start a
program e.g. via an external pulse.
Four inputs are available.