User Manual
EN
r
esistance
Measuring range
Resolution
Accuracy
600 Ω
0.1 Ω
±(1,3% + 5d)
6 kΩ
1 Ω
±(1,3% + 5d)
60 kΩ
10 Ω
±(1,0% + 5d)
600 kΩ
100 Ω
±(1,0% + 5d)
6 MΩ
1 kΩ
±(1,0% + 5d)
60 MΩ
10 kΩ
±(1,5% + 5d)
a
ccUracy
Measuring range
Resolution
Accuracy
6 nF
1 pF
± (3.0% + 5d)
60 nF
10 pF
± (3.0% + 5d)
600 nF
100 pF
± (3.0% + 5d)
6 µF
1 nF
± (3.0% + 5d)
60 µF
10 nF
± (3.0% + 5d)
600 µF
100 nF
± (3.5% + 5d)
6 mF
1 µF
± (5.0% + 6d)
60 mF
10 µF
± (10.0% + 8d)
f
reqUency
Measuring range
Accuracy
10 Hz ~ 100 kHz
± (1.0%+5d)
1 MHz
± (3.0%+5d)
20 MHz
± (4.0%+10d)
c
ontrols
MP 25700
1
2
3
VC803
1
Power button:
Briefly
press this button to turn
on and the device will
automatically start in
voltage / resistance auto
detection mode. (Voltage,
resistance, or current can
be measured directly.) To
turn off, press and hold
this button for at least
2 seconds.
2
„H” button:
Press once
to keep the current value
on the display. „
H
” appears
on the display. To clear
the value, press the
button again.
Backlight / Flashlight:
Press and hold for at least
2 seconds to backlight
the display and turn on
the flashlight function. To
deactivate these functions, press and hold the button again.
3
Function toggle button:
Briefly press the button once
and select (NCV) or (CAP) or (Hz). Press and hold this button
for approx. 2 seconds to exit the function mode and enter
the voltage / resistance auto detection mode.
U
sing
tHe
MUltiMeter
DC / AC voltage measurement (> 1 V)
1. Measurement is only possible if the voltage is greater
than 1 V.
2. Connect the red instrument cable to the appropriate
socket, then connect the black instrument cable to the
„
COM
” socket.
3. DC or AC voltage measurement is automatic.
4. Touch the probes to the appropriate point on the test
circuit to measure the voltage.
5. Read the measured voltage from the display.
Resistance measurement
1. Connect the red wire to the appropriate socket, then
connect the black instrument cable to the „
COM
” socket.
2. The resistance is measured automatically.
3. Touch the probes to the appropriate point on the test
circuit to measure the resistance of the circuit.
4. Read the measured resistance from the display.
Continuity test
1. Connect the red instrument cable to the appropriate
socket, then connect the black instrument cable to the
„
COM
” socket.
2. The continuity test measurement is performed
automatically.