![Maxwell 25700 User Manual Download Page 3](http://html.mh-extra.com/html/maxwell/25700/25700_user-manual_1746502003.webp)
User Manual
EN
p
arts
of
display
AUTO
OUT
hFE
MIN MAX
%
AVnF
m
μ
Mk Hz
Ω
AUTO
– Automatic measuring
– High voltage warning
– Dangerous voltage warning
hFE
– Transistor testing
– Diode test
– Continuity test
– Battery charge
MIN / MAX
- Measurement range
– Data hold
– DC current sign
– AC current sign
– Rectangular impulse
Unit of measurement (
%,°C, °F, MK, Ohm, Hz, m, AV, nF
)
g
eneral
tecHnical
paraMeters
Display
4 digits, LCD
Maximum characteristic
6000
Range
Automatic
Sampling
About 3x in one second
Low power indication
True RMS
Data hold
Backlight
Flashlight
Automatic shutdown
Operating temperature
0 ºC - 40 ºC, <75%
humidity
Storage temperature
-20 ºC - 60 ºC, <85%
humidity
Dimensions
187 x 66 x 32 mm
Continuity
NCV
s
pecification
Specification
Accuracy measurements were measured one year after
calibration at a temperature between 18 ºC and 28 ºC and
a relative humidity of <75%. Format: ± (% measured value
+ digit value)
ac v
oltage
Measuring range
Resolution
Accuracy
6 V
0.001 V
± (1%+5d)
60 V
0.01 V
± (1%+5d)
600 V
0.1 V
± (1%+8d)
Frekvencia
40 Hz ~ 1 kHz
Note: The device malfunctions, burning of
not measure voltages above 600 V in order to
prevent!
dc v
oltage
Measuring range
Resolution
Accuracy
6 V
0.001 V
± (0.5%+5d)
60 V
0.01 V
± (0.5%+5d)
600 V
0.1 V
± (0.8%+5d)
Note: The device malfunctions, burning of
not measure voltages above 600 V in order to
prevent!
ac c
Urrent
Measuring range
Resolution
Accuracy
1 A
1 mA
± (1.2%+6d)
10 A
10 mA
± (1.2%+6d)
Frequency range
40 Hz ~ 1 kHz
dc c
Urrent
Measuring range
Resolution
Accuracy
1 A
1 mA
± (1.0%+5d)
10 A
10 mA
± (1.0%+5d)
Frequency range 40 Hz ~ 1 kHz