Maxtec Handi+N2 Operating Manual & Instructions For Use Download Page 9

866.4.Maxtec    www.maxtecinc.com

3

For best results, perform the calibration procedure at a temperature close to

the temperature where analysis will occur.

Allow adequate time for the sensor to equilibrate to a new ambient temperature.

CAUTION: 

“CAL Err St” may result from a sensor that has not reached thermal 

equilibrium.

Pressure Effects

Readings from the N2 analyzer are proportional to the partial pressure of 
oxygen. The partial pressure is equal to the concentration times the absolute 
pressure. Thus, the readings are proportional to the concentration if the 
pressure is held constant. Therefore, the following are recommended:

Calibrate the N2 analyzer at the same pressure as the sample gas.

If sample gases flow through tubing, use the same apparatus and flow rates

when calibrating as when measuring.

The N2 analyzer oxygen sensor has been tested at pressures up to two

atmospheres absolute. Calibration or operation above this pressure is beyond
the intended use.

Humidity Effects

Humidity (non-condensing) has no effect on the performance of the N2 
analyzer other than diluting the gas, as long as there is no condensation. 
Depending on the humidity, the gas may be diluted by as much as 4%, which 
proportionally reduces the oxygen concentration. The device responds to the 
actual oxygen concentration rather than the dry concentration. Environments 
where condensation may occur are to be avoided since moisture may obstruct 
passage of gas to the sensing surface, resulting in erroneous readings and 
slower response time. For this reason, the following is recommended:

Avoid usage in environments greater than 95% relative humidity.

         866.4.Maxtec   www.maxtec.com

Summary of Contents for Handi+N2

Page 1: ...E Handi N2 O P E R A T I N G M A N U A L I N S T R U C T I O N S F O R U S E R 2 1 8 M 0 6 I n d u s t r i a l R218M06 Rev E N2...

Page 2: ......

Page 3: ...th our operating instructions Based on our product evaluation our sole obligation under the foregoing warranty is limited to making replacements repairs or issuing credit for equipment found to be def...

Page 4: ...he oxygen sensor is a sealed device containing a mild acid electrolyte lead Pb and lead acetate Lead and lead acetate are hazardous waste constituents and should be disposed of properly or returned fo...

Page 5: ...DE IV 1 0 INTRODUCTION 1 2 0 N2 ANALYZER FEATURES AND FUNCTIONS 1 3 0 PRE USE CHECKOUT CALIBRATION 2 4 0 OPERATION PRINCIPALS 2 5 0 FACTORS INFLUENCING ACCURATE READINGS 2 6 0 CALIBRATION ERRORS AND E...

Page 6: ...he Handi Do not throw away Follow local guidelines for disposal Follow instructions for use Do Not On off Button Warnings Calibration Button Meets ETL standards LOT IPX1 Lot code Batch code Drip Proof...

Page 7: ...2 Analyzer enters its sleep power off mode The N2 Analyzer will automatically enter the sleep mode after 2 minutes from the last time the unit was energized You can manually turn off the analyzer by p...

Page 8: ...ncing accurate readings 5 The N2 Analyzer is ready to use 4 0 OPERATION PRINCIPALS The instrument display corresponds directly to the oxygen sensor The oxygen diffuses through the membrane and an elec...

Page 9: ...ng use the same apparatus and flow rates when calibrating as when measuring The N2 analyzer oxygen sensor has been tested at pressures up to two atmospheres absolute Calibration or operation above thi...

Page 10: ...calibration If unit repeats this error more than three times contact Maxtec Customer Service for possible sensor replacement CAL Err hi Sensor voltage too high Press and hold the Calibration Button fo...

Page 11: ...or 90 step change at 25 C Warm up Time None required Operating Temperature 15 C 40 C 59 F 104 F Storage Temperature 15 C 50 C 5 F 122 F Operating Pressure Atmospheric pressure to 3psig Environmental G...

Reviews: