Maxtec Handi+N2 Operating Manual & Instructions For Use Download Page 10

866.4.Maxtec    www.maxtecinc.com

4

 6.0 CALIBRATION ERRORS AND ERROR CODES

The N2 analyzers have a self test feature built into the software to detect faulty 
calibrations, oxygen sensor failures, and low operating voltage. These are listed 
below, and include possible actions to take, if an error code occurs. 

E03: No valid calibration data available

Make sure unit has reached thermal equilibrium. Press and hold the 
Calibration Button     for three seconds to manually force a new calibration. 

E04: Battery below minimum operating voltage

Unit is at end of life, see page I for proper disposal.

CAL Err St: O2 Sensor reading not stable

Wait for displayed oxygen reading to stabilize when calibrating the device at 
100% oxygen. Wait for unit to reach thermal equilibrium (Please note that this 
can take up to one half hour, if the device is stored in temperatures outside the 
specified operating temperature range). 

CAL Err lo: Sensor voltage too low

Press and hold the Calibration Button     for three seconds to manually force 
a new calibration. If unit repeats this error more than three times, contact 
Maxtec Customer Service for possible sensor replacement.

CAL Err hi: Sensor voltage too high

Press and hold the Calibration Button     for three seconds to manually force 
a new calibration. If unit repeats this error more than three times, contact 
MAxtec Customer Service for possible sensor replacement.

CAL Err Bat: Battery voltage too low to recalibrate

Unit is at end of life, see page I for proper disposal. 

 7.0 CLEANING, MAINTENANCE, AND DISPOSAL

Store the N2 analyzer in a temperature similar to its ambient environment of 
daily use. The instructions given below describes the methods to clean the 
instrument, sensor and its accessories: 

Instrument

When cleaning or disinfecting the exterior of the N2 analyzer, take

appropriate care to prevent any solution from entering the instrument.

 Do not immerse unit in fluids.

Oxygen Sensor

Clean the sensor with a cloth moistened with a 65% alcohol / water solution.

We do not recommend use of spray disinfectants because they can contain

salt, which can accumulate in the sensor membrane and impair readings.

Do not throw away. Dispose of properly in accordance with local regulations.

         866.4.Maxtec   www.maxtec.com

Summary of Contents for Handi+N2

Page 1: ...E Handi N2 O P E R A T I N G M A N U A L I N S T R U C T I O N S F O R U S E R 2 1 8 M 0 6 I n d u s t r i a l R218M06 Rev E N2...

Page 2: ......

Page 3: ...th our operating instructions Based on our product evaluation our sole obligation under the foregoing warranty is limited to making replacements repairs or issuing credit for equipment found to be def...

Page 4: ...he oxygen sensor is a sealed device containing a mild acid electrolyte lead Pb and lead acetate Lead and lead acetate are hazardous waste constituents and should be disposed of properly or returned fo...

Page 5: ...DE IV 1 0 INTRODUCTION 1 2 0 N2 ANALYZER FEATURES AND FUNCTIONS 1 3 0 PRE USE CHECKOUT CALIBRATION 2 4 0 OPERATION PRINCIPALS 2 5 0 FACTORS INFLUENCING ACCURATE READINGS 2 6 0 CALIBRATION ERRORS AND E...

Page 6: ...he Handi Do not throw away Follow local guidelines for disposal Follow instructions for use Do Not On off Button Warnings Calibration Button Meets ETL standards LOT IPX1 Lot code Batch code Drip Proof...

Page 7: ...2 Analyzer enters its sleep power off mode The N2 Analyzer will automatically enter the sleep mode after 2 minutes from the last time the unit was energized You can manually turn off the analyzer by p...

Page 8: ...ncing accurate readings 5 The N2 Analyzer is ready to use 4 0 OPERATION PRINCIPALS The instrument display corresponds directly to the oxygen sensor The oxygen diffuses through the membrane and an elec...

Page 9: ...ng use the same apparatus and flow rates when calibrating as when measuring The N2 analyzer oxygen sensor has been tested at pressures up to two atmospheres absolute Calibration or operation above thi...

Page 10: ...calibration If unit repeats this error more than three times contact Maxtec Customer Service for possible sensor replacement CAL Err hi Sensor voltage too high Press and hold the Calibration Button fo...

Page 11: ...or 90 step change at 25 C Warm up Time None required Operating Temperature 15 C 40 C 59 F 104 F Storage Temperature 15 C 50 C 5 F 122 F Operating Pressure Atmospheric pressure to 3psig Environmental G...

Reviews: