Maxtec Handi+N2 Operating Manual & Instructions For Use Download Page 8

866.4.Maxtec    www.maxtecinc.com

2

 3.0 PRE-USE CHECKOUT/CALIBRATION

Follow these steps before using the N2 Analyzer
1. Prior to turning on the unit, a protective film covering the threaded sensor
face must be removed. After removing the film, wait approximately 20 minutes
for the sensor to reach equilibrium.
2. Pre-assembly, if required.

• 

Thread the barbed adapter onto the oxygen sensor.

Connect the clear tubing to the barbed adapter.

3. Using the "ON/OFF" key      , make sure the unit is in the power "ON” mode.
4. Press and hold the Calibration Key      for 3 seconds until the display
reads “CAL”. This will calibrate the N2 Analyzer to room air. Thereafter, we
recommend calibration on a weekly basis.
A new calibration is required when:

The measured N2 percentage in 79.1% N2 is above 80.1% N2

The measured N2 percentage in 79.1% N2 is below 78.1% N2

If you are unsure about the displayed N2 percentage. (see Factors

influencing accurate readings.)

5. The N2 Analyzer is ready to use.

 4.0 OPERATION PRINCIPALS

The instrument display corresponds directly to the oxygen sensor. The oxygen 
diffuses through the membrane and an electrical current is generated that is 
proportional to the partial pressure of oxygen in the gas sample. The oxygen 
percentage is subtracted from 100, with the remainder being displayed as 
percent nitrogen. The sensor has a minimal response to gases other than oxygen.

 5.0 FACTORS INFLUENCING ACCURATE READINGS

Elevation Changes

Changes in elevation result in a reading error of approximately 1% of reading

per 250 feet.

In general, calibration of the instrument should be performed when elevation

at which the product is being used changes by more than 500 feet.

Temperature Effects

The N2 analyzer will hold calibration and read correctly within ±3% when in 
thermal equilibrium within the operating temperature range. The device must 
be thermally stable when calibrated and allowed to thermally stabilize after 
experiencing temperature changes before readings are accurate. 

         866.4.Maxtec   www.maxtec.com

Summary of Contents for Handi+N2

Page 1: ...E Handi N2 O P E R A T I N G M A N U A L I N S T R U C T I O N S F O R U S E R 2 1 8 M 0 6 I n d u s t r i a l R218M06 Rev E N2...

Page 2: ......

Page 3: ...th our operating instructions Based on our product evaluation our sole obligation under the foregoing warranty is limited to making replacements repairs or issuing credit for equipment found to be def...

Page 4: ...he oxygen sensor is a sealed device containing a mild acid electrolyte lead Pb and lead acetate Lead and lead acetate are hazardous waste constituents and should be disposed of properly or returned fo...

Page 5: ...DE IV 1 0 INTRODUCTION 1 2 0 N2 ANALYZER FEATURES AND FUNCTIONS 1 3 0 PRE USE CHECKOUT CALIBRATION 2 4 0 OPERATION PRINCIPALS 2 5 0 FACTORS INFLUENCING ACCURATE READINGS 2 6 0 CALIBRATION ERRORS AND E...

Page 6: ...he Handi Do not throw away Follow local guidelines for disposal Follow instructions for use Do Not On off Button Warnings Calibration Button Meets ETL standards LOT IPX1 Lot code Batch code Drip Proof...

Page 7: ...2 Analyzer enters its sleep power off mode The N2 Analyzer will automatically enter the sleep mode after 2 minutes from the last time the unit was energized You can manually turn off the analyzer by p...

Page 8: ...ncing accurate readings 5 The N2 Analyzer is ready to use 4 0 OPERATION PRINCIPALS The instrument display corresponds directly to the oxygen sensor The oxygen diffuses through the membrane and an elec...

Page 9: ...ng use the same apparatus and flow rates when calibrating as when measuring The N2 analyzer oxygen sensor has been tested at pressures up to two atmospheres absolute Calibration or operation above thi...

Page 10: ...calibration If unit repeats this error more than three times contact Maxtec Customer Service for possible sensor replacement CAL Err hi Sensor voltage too high Press and hold the Calibration Button fo...

Page 11: ...or 90 step change at 25 C Warm up Time None required Operating Temperature 15 C 40 C 59 F 104 F Storage Temperature 15 C 50 C 5 F 122 F Operating Pressure Atmospheric pressure to 3psig Environmental G...

Reviews: