Maxtec Handi+N2 Operating Manual & Instructions For Use Download Page 4

866.4.Maxtec    www.maxtecinc.com

II

 WARNINGS

 

Indicates a potentially hazardous situation, if not avoided, could result in 

death or serious injury.

» Never allow an excess length of tubing, or any accessory near a person’s head or

neck, which may result in strangulation.

»  Before use, all individuals who will be using the N2 Analyzer must become thoroughly

familiar with the information contained in this Operation Guide.

» Strict adherence to the operating instructions is necessary for safe, effective product

performance. This product will perform only as designed if installed and operated in
accordance with the manufacturer’s operating instructions.

» Use only genuine accessories and replacement parts. Failure to do so may seriously

impair the analyzer’s performance.

» Repair or alteration of the N2 Analyzer beyond the scope of the maintenance

instructions, or by anyone other than an authorized service person, could cause the
product to fail to perform as designed.

» Calibrate the N2 analyzer weekly when in operation, or if environmental conditions

change significantly. (ie, Elevation, Temperature, Pressure, Humidity — refer to
“Factors Influencing Accurate Readings”).

» Use of the N2 analyzer near devices that generate electrical fields may cause erratic readings
» Ensure proper tire inflation pressure after use, if required.
» The oxygen sensor is a sealed device containing a mild acid electrolyte, lead (Pb),

and lead acetate. Lead and lead acetate are hazardous waste constituents and
should be disposed of properly, or returned for proper disposal or recovery.

» Dropping the device can adversely affect its performance.

Do not immerse the device in any cleaning solution, autoclave or expose the sensor

to high temperatures (> 70°C).

Do not over pressurize the sensor. Doing so may destroy the sensor and void the

warranty. To avoid over pressurization only allow 3 psi (or 2 liters per minute) of

gas to come in contact with the sensor membrane.

CAUTION:

Indicates a potentially hazardous situation, if not avoided, could result in

minor or moderate injury and property damage.

» Read the manual in its entirety before attempting use.

»  Always use protective eyewear and observe proper safety procedures when working

with pressurized gases.

»  Dispose of the N2 properly when it has expired.

» Ensure the protective freshness seal has been removed from the sensing port before use.

» Ensure the N2 has been properly calibrated before use.

» If the N2 display goes blank immediately after the on button is pushed, or the

N2 will not properly calibrate, the unit has expired.

Do not use, dispose of properly.

» The display is not valid when in Over Range Mode. Recalibrate the N2 and

observe the proper operating procedure.

» Never immerse the N2 or expose it to high humidity or moisture. It is not watertight.

» Never expose the N2 to high temperatures.

         866.4.Maxtec   www.maxtec.com

!

Summary of Contents for Handi+N2

Page 1: ...E Handi N2 O P E R A T I N G M A N U A L I N S T R U C T I O N S F O R U S E R 2 1 8 M 0 6 I n d u s t r i a l R218M06 Rev E N2...

Page 2: ......

Page 3: ...th our operating instructions Based on our product evaluation our sole obligation under the foregoing warranty is limited to making replacements repairs or issuing credit for equipment found to be def...

Page 4: ...he oxygen sensor is a sealed device containing a mild acid electrolyte lead Pb and lead acetate Lead and lead acetate are hazardous waste constituents and should be disposed of properly or returned fo...

Page 5: ...DE IV 1 0 INTRODUCTION 1 2 0 N2 ANALYZER FEATURES AND FUNCTIONS 1 3 0 PRE USE CHECKOUT CALIBRATION 2 4 0 OPERATION PRINCIPALS 2 5 0 FACTORS INFLUENCING ACCURATE READINGS 2 6 0 CALIBRATION ERRORS AND E...

Page 6: ...he Handi Do not throw away Follow local guidelines for disposal Follow instructions for use Do Not On off Button Warnings Calibration Button Meets ETL standards LOT IPX1 Lot code Batch code Drip Proof...

Page 7: ...2 Analyzer enters its sleep power off mode The N2 Analyzer will automatically enter the sleep mode after 2 minutes from the last time the unit was energized You can manually turn off the analyzer by p...

Page 8: ...ncing accurate readings 5 The N2 Analyzer is ready to use 4 0 OPERATION PRINCIPALS The instrument display corresponds directly to the oxygen sensor The oxygen diffuses through the membrane and an elec...

Page 9: ...ng use the same apparatus and flow rates when calibrating as when measuring The N2 analyzer oxygen sensor has been tested at pressures up to two atmospheres absolute Calibration or operation above thi...

Page 10: ...calibration If unit repeats this error more than three times contact Maxtec Customer Service for possible sensor replacement CAL Err hi Sensor voltage too high Press and hold the Calibration Button fo...

Page 11: ...or 90 step change at 25 C Warm up Time None required Operating Temperature 15 C 40 C 59 F 104 F Storage Temperature 15 C 50 C 5 F 122 F Operating Pressure Atmospheric pressure to 3psig Environmental G...

Reviews: