Maxtec Handi+N2 Operating Manual & Instructions For Use Download Page 7

866.4.Maxtec    www.maxtecinc.com

1

Follow instructions for use

Do Not

Calibration Button

Lot code/Batch code

Drip Proof

Serial Number

Catalog Number

 1.0 INTRODUCTION

This manual describes the function; operation and maintenance of the N2 
Analyzer. The intent of this manual is to describe the function of the N2 
Analyzer only. The final assembly manufacturer should provide operating 
instructions for the completed assembly. The N2 analyzer is engineered for 
long life, maximum reliability and stable performance.

NOTE: In order to obtain optimum performance from your analyzer, all 

operation and maintenance must be performed in accordance with this 

manual. Please read the manual thoroughly before using the analyzer and 

do not attempt any repair or procedure that is not described herein. We 

cannot warranty any damage resulting from misuse, unauthorized repair or 

improper maintenance of the instrument.

 2.0 N2 ANALYZER FEATURES AND FUNCTIONS

LCD Display: 

A 3-digit display provides a direct readout of nitrogen 

concentration in the range of 0 - 99.9%. The display is blank when the N2 
Analyzer enters its sleep (power off) mode. The N2 Analyzer will automatically 
enter the sleep mode after 2 minutes from the last time the unit was energized. 
You can manually turn off the analyzer by pressing the on/off switch.

ON Button/Auto OFF: 

Use this button to turn the N2 Analyzer on or off. When 

the N2 Analyzer is in the Sleep (power off) mode, the LCD display is blank. 
When the ON button is pressed once, the analyzer will display the nitrogen 
concentration for 2 minutes. Pressing the ON button during this 2 minute 
"window" will prolong the ON period to 2 minutes from the most recent time 
that the button was pushed.

Over Range Indicator: 

The appearance of a decimal point after the first digit 

means that the N2 Analyzer is reading in excess of 99.9%.
Example: 0.0.0 = 100% 0.0.1 = 101%
0.0.2 = 102% (etc).

Calibration Key:

 This key is used to calibrate the device. Holding the key for 

more than three seconds will force the device to enter a calibration mode.

Oxygen Sensor:

 This is used to measure oxygen concentration in sample gas.

Sample Inlet Connection:

 This is the port at which the device is connected to 

determine nitrogen concentration.

         866.4.Maxtec   www.maxtec.com

Summary of Contents for Handi+N2

Page 1: ...E Handi N2 O P E R A T I N G M A N U A L I N S T R U C T I O N S F O R U S E R 2 1 8 M 0 6 I n d u s t r i a l R218M06 Rev E N2...

Page 2: ......

Page 3: ...th our operating instructions Based on our product evaluation our sole obligation under the foregoing warranty is limited to making replacements repairs or issuing credit for equipment found to be def...

Page 4: ...he oxygen sensor is a sealed device containing a mild acid electrolyte lead Pb and lead acetate Lead and lead acetate are hazardous waste constituents and should be disposed of properly or returned fo...

Page 5: ...DE IV 1 0 INTRODUCTION 1 2 0 N2 ANALYZER FEATURES AND FUNCTIONS 1 3 0 PRE USE CHECKOUT CALIBRATION 2 4 0 OPERATION PRINCIPALS 2 5 0 FACTORS INFLUENCING ACCURATE READINGS 2 6 0 CALIBRATION ERRORS AND E...

Page 6: ...he Handi Do not throw away Follow local guidelines for disposal Follow instructions for use Do Not On off Button Warnings Calibration Button Meets ETL standards LOT IPX1 Lot code Batch code Drip Proof...

Page 7: ...2 Analyzer enters its sleep power off mode The N2 Analyzer will automatically enter the sleep mode after 2 minutes from the last time the unit was energized You can manually turn off the analyzer by p...

Page 8: ...ncing accurate readings 5 The N2 Analyzer is ready to use 4 0 OPERATION PRINCIPALS The instrument display corresponds directly to the oxygen sensor The oxygen diffuses through the membrane and an elec...

Page 9: ...ng use the same apparatus and flow rates when calibrating as when measuring The N2 analyzer oxygen sensor has been tested at pressures up to two atmospheres absolute Calibration or operation above thi...

Page 10: ...calibration If unit repeats this error more than three times contact Maxtec Customer Service for possible sensor replacement CAL Err hi Sensor voltage too high Press and hold the Calibration Button fo...

Page 11: ...or 90 step change at 25 C Warm up Time None required Operating Temperature 15 C 40 C 59 F 104 F Storage Temperature 15 C 50 C 5 F 122 F Operating Pressure Atmospheric pressure to 3psig Environmental G...

Reviews: