DBF01O04.docx
(MAVIG #7111730)
page 20 of 22
3.5.
F
ASTENING OF AN
A
SSEMBLY
A
ID
MAVIG supplies an
eye bolt
that can be fastened to the
ceiling plate
.
It will be used as an anchor point for lifting e.g. monitors and the GD60D- suspension
system.
Apply Loctite 270
to the bolt.
Screw the eye bolt with the washer to one of the four M12 threads in the ceiling
plate.
Tighten the bolt with the force of your hand only.
4.
S
AFE
D
ISPOSAL
The system/device must be disposed of in accordance with the nationally applicable
provisions at an appropriate recycling facility for electrical and electronic equipment. You
may also return the system/device to MAVIG for disposal. If you have any questions,
please contact us at: e-Mail:
Tel.:
+49 (0) 89 / 420 96 0
RoHS Conformity:
The support arm system meets the requirements of Directive 2011/65/EU, RohS
(Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic
Equipment) established for device group 9.
WARNING: falling parts / injury / death!
DON’T apply torque with a tool to the eye bolt!
Tighten the bolt with the force of your hand only.