3.2.
F
ASTENING THE
C
EILING
P
LATE
NOTE:
The diameter of the opening in the false ceiling should be not
wider than 620mm.
Otherwise, the opening will not be covered with the GD60 ceiling
covers. For dimensions, see below.
Fig. 7
Fix the ceiling plate to the ceiling utilizing the six holes with diameter 17mm at 360mm x
451mm.
The ceiling plate can be fastened to the ceiling using six suitable screws. These screws
must be dimensioned according to the conditions of the ceiling and provided by the
customer.
Make sure that the plate is sitting flush to the ceiling in order to ensure optimum load
distribution.
This can be checked by use of a spirit level (accuracy: 1mm/m).
We recommend to use a digital spirit level.