SPECIFICATIONS
LNC MP Patient Cables are indicated for use with the following devices and sensors:
When used with Masimo SET® Technology:
• Masimo SET Intellivue Module
• Masimo SET Intellivue Multi-Measurement Server (MMS)
• Masimo SET Intellivue Multi-Measurement Server X2 (MMS X2)
Sensors
Body Weight
Saturation Accuracy
(70–100% SpO
2
)
Pulse Rate Accuracy
(25–240 bpm)
Low Perfusion Accuracy
(70–100% SpO
2
)
No Motion
Motion
No Motion
Motion
Saturation
Pulse Rate
LNCS Adtx/Adtx-3
> 30 kg
2%
3%
3 bpm
5 bpm
2%
3 bpm
LNCS Pdtx/Pdtx-3
10–50 kg
2%
3%
3 bpm
5 bpm
2%
3 bpm
LNCS Inf/Inf-3
3–20 kg
2%
3%
3 bpm
5 bpm
2%
3 bpm
LNCS Neo/Neo-3/Neo-L
< 3 kg
3%
3%
3 bpm
5 bpm
3%
3 bpm
> 40 kg
2%
3%
3 bpm
5 bpm
2%
3 bpm
LNCS NeoPt/NeoPt-3/NeoPt-L
< 1 kg
3%
3%
3 bpm
5 bpm
3%
3 bpm
LNCS Trauma
> 30 kg
2%
3%
3 bpm
5 bpm
2%
3 bpm
LNCS Newborn Neonatal
< 3 kg
3%
3%
3 bpm
5 bpm
3%
3 bpm
LNCS Newborn Infant/
Pediatric
3–10 kg
2%
3%
3 bpm
5 bpm
2%
3 bpm
10–30 kg
2%
3%
3 bpm
5 bpm
2%
3 bpm
LNCS DCI
> 30 kg
2%
3%
3 bpm
5 bpm
2%
3 bpm
LNCS DCIP
10–50 kg
2%
3%
3 bpm
5 bpm
2%
3 bpm
LNCS TC-I
> 30 kg
3.5%
N/A
3 bpm
N/A
3.5%
3 bpm
LNCS TF-I®
> 30 kg
2%
N/A
3 bpm
N/A
2%
3 bpm
LNCS YI
1–3 kg
3%
3%
3 bpm
5 bpm
3%
3 bpm
> 3 kg
2%
N/A
N/A
N/A
2%
N/A
LNCS TF-I
> 30 kg
2%
N/A
3 bpm
N/A
N/A
N/A
LNCS DBI
> 30 kg
2%
N/A
3 bpm
N/A
2%
3 bpm
LNCS S-ROS 3U
universal
See applicable S-DOS sensor below for accuracy specifications when used in combination
with S-ROS optical sensor.
Resposable S-DOS 25
> 30 kg
2%
3%
3 bpm
5 bpm
2%
3 bpm
Resposable S-DOS 20
10–50 kg
2%
3%
3 bpm
5 bpm
2%
3 bpm
Resposable S-DOS 25L
< 3 kg
3%
3%
3 bpm
5 bpm
3%
3 bpm
> 30 kg
2%
3%
3 bpm
5 bpm
2%
3 bpm
Resposable S-DOS 20L
3–10 kg
2%
3%
3 bpm
5 bpm
2%
3 bpm
10–30 kg
2%
3%
3 bpm
5 bpm
2%
3 bpm
Resposable S-DOS 30L
3–10 kg
2%
3%
3 bpm
5 bpm
2%
3 bpm
> 40 kg
2%
3%
3 bpm
5 bpm
2%
3 bpm
Resposable S-DOS 25Pt-L
< 1 kg
3%
3%
3 bpm
5 bpm
3%
3 bpm
Resposable S-DOS 20L-500
3–10 kg
2%
3%
3 bpm
5 bpm
2%
3 bpm
10–30 kg
2%
3%
3 bpm
5 bpm
2%
3 bpm
A
rms
accuracy is a statistical calculation of the difference between device measurements and reference measurements. Approximately two-thirds
of the device measurements fell /-
A
rms
of the reference measurements in a controlled study.
The LNCS TF-I and TC-I sensors were not validated under motion conditions.