User Manual
P a g e
|
95
of
100
6.6 Electrical Measurement Parameters and Accuracy
Parameter Range
Resolution Accuracy
V
0-30.00 VDC
0.001 V
+0.02% of r 2 count
mA
0-24.00 mA
0.001 mA
+0.02% of r 2 count
6.7 Electrical Simulation Parameters and Accuracy
Parameter
Range
Resolution
Accuracy
V
0-12.00 VDC
0.001 V
+0.02% of r 2 count
mA
0-24.00 mA
0.001 mA
+0.02% of r 2 count
6.8 Resistance Measurement 0 ... 4000 ohm
Range
Resolution
Accuracy
0….1600
Ω
10 m
Ω
4 Wire Connection:
0.02% of r 0.01
Ω
3 Wire Connection:
0.02% of r 0.015
Ω
1600….4000
Ω
100 m
Ω
4 Wire Connection:
0.02% of r 0.1
Ω
3 Wire Connection:
0.02% of r 0.15
Ω
6.9 Resistance Simulation 0 ... 4000 ohm
Range
Resolution
Accuracy*
0….400
Ω
10 m
Ω
0.02% of r 0.02
Ω
400….4000
Ω
100 m
Ω
0.02% of r 0.15
Ω
6.10 RTD Measurement and Simulation
1)
Pt10 …… Pt1000, -200 …… 850 ºC
Range
Resolution
Accuracy
-
200….200 °C
Pt10…Pt400:
0.01
°C
4 wire Measurement: 0.15°C
Simulation*: 0.15 °C
200….600 °C
4 wire Measurement: 0.2 °C