Procedure for calibration of Frequency Source
To enter into the Frequency Source Calibration, Select the
Frequency
option in Calibration page and Source Frequency option in CAL.
Frequency page.
All Frequency input are divided into 2 groups.
Frequency SOURCE Group
Recommended
Point
1
100.000 Hz
2
5000.0 Hz
Example: - Calibrating Group-1
Select Group 1 in Freq. Group ListBox.
For calibration,
Select
Measured
Edit Box, when this Edit Box is selected UC-12
will source value that is seen in
Expected
Value (here UC-12 will
source 100 Hz).
Now Measure the source value in Reliable Measure Unit. For
Example the external measure unit is measuring 100.020 Hz. Then
enter 100.020 Hz value in
Measured
Edit Box & Press
MENU/ENT
key to calibrate the SPAN.
Follow the same procedure to calibrate the other Frequency Source
Group.
Note:
Usually, UC-12 calibrates using the above steps in one/two time only.
But if the Input / Output are very out from the desired value repeat the
Calibration steps until the device calibrated properly.