Procedure for calibration of Ambient Temperature
To enter into the Ambient Temperature Calibration, Select the
TC
option in Calibration page and Ambient Cal. option in CAL. TC page.
Example: - Calibrating CJ Temperature
CJ measured value shows the current measured temperature of the
TC Terminals in the Display Unit selected in TC Setup Page. For
example if it shows 25.0 °C and the real temperature of the TC
terminal is 25.5 °C then enter 25.5 in Actual Editbox to calibrate CJ
Temperature.
Note:
Please make sure that your unit is in proper Input TC Type and CJC is
in AUTO mode.