Example: - Calibrating mA Output
To calibrate
ZERO for group 1,
Select
ZERO Measured
Edit Box, when this Edit Box is selected
UC-12 will source value that is seen in
ZERO Expected
Value (here
UC-12 will source 1.000mA).
Now Measure the source value in Reliable Measure Unit. For
Example the external measure unit is measuring 1.020mA. Then
enter 1.020 value in
ZERO Measured
Edit Box & Press
MENU/ENT
key to calibrate the ZERO.
To calibrate
SPAN for group 1,
Select
SPAN Measured
Edit Box, when this Edit Box is selected
UC-12 will source value that is seen in
SPAN Expected
Value (here
UC-12 will source 10.000mA).
Now Measure the source value in Reliable Measure Unit. For
Example the external measure unit is measuring 10.020mA. Then
enter 10.020 value in
SPAN Measured
Edit Box & Press
MENU/ENT
key to calibrate the SPAN.
Follow the same procedure to calibrate group 2.
The same procedure is applicable for other output Type.
Note:
No need to separately calibrate mA (2W) Output.