Ref. No: m94A/OM/201 
Issue No: 01

 

 

Operator’s manual 

Page 21 of 26 

(6) 

LCHk

 (Link Check Timer) Mode: 

 

LCHk 

is for ensuring communication link between HT16Ew & CDU/85xx+ remains 

healthy

. User can set the 

LCHk 

minute value with INC key. Device will transmits frame 

at regular interval which is defined by 

LCHk 

value. At every Link-Check Timer HT16Ew 

transmits one communication frame to CDU/85xx+ irrespective of 

PTMR

 value& 

exception. User can only configured 

LCHk 

value from below set: 

{1,5,10,15,30,60,120,180,240,360,480,720,1440,2880} this values are in minutes. 

 

Factory default 

LCHk 

value is 

0060 

minutes.

 

 

 

8.5 

EXCP

(Exception menu):

 

It let the user to configure exception protocol parameters. Exception is defined as pre-
defined  band  of  temperature  and  humidity.  Using  this  band  transmitter  will  only 
transmit the  frame  if  value  of temperature  or  humidity crosses the  band  at  particular 
PTMR interval. If end user set this band according to his  requirement, the battery life 
will  increase  according  and  this  will  be  the  benefit  for  end  user.  This  menu  further 
divided in sub categories below,

 

(1) 

Tbnd

 (Temperature Band) Mode: 

  

Tbnd mode used to configure temperature band for exception protocol of transmitter 
unit. Suppose Tbnd is set to 0.5then frame will only transmits when current 
temperature deviate ±0.5 °C from last measured temperature. 

On the press of MENU key, screen will appear which will show the temperature band 
for exception protocol. To change the date user first press MENU key, the first digit will 
blink. Now using INC key user can change first digit. Now if user press the DCR key, 
cursor moves to second digit. Now using INC key user can change second digit. 

On the press of MENU key again, save Yes/ save No (

S.Y

S.n

 ) screen will appear for 

saving new change. If user will press “

S.n

” (INC Key) then it will discard the new 

changes. If user will press “

S.Y

” (DCR Key) then it will set unit as day.

 

Factory default value of  

tbnd 

value is 

0.5 

°C

.

 

If user press DCR key in between, it will shows the main menu screen. 

 

(2) 

Hbnd

 (Humidity Band) Mode: 

 

Hbnd

 mode used to configure Humidity band for exception protocol of transmitter unit. 

Suppose 

Hbnd

 is set to

2.0

 then frame will only transmits when current measured 

humidity deviate 

±

2.0 

%

RH 

from last measured humidity.

 

On the press of MENU key, screen will appear which will show the Humidity band for 
exception protocol. To change the date user first press MENU key, the first digit will 

Summary of Contents for HT16Ew

Page 1: ...26 User Manual HT16Ew WIRELESS RH T TRANSMITTER DOC m94A OM 201Issue No 01 Copyright 2018 Masibus Automation and Instrumentation Pvt Ltd Customer Support support masibus com Toll Free Number India On...

Page 2: ...MOUNTING 7 7 1 Installation with Back plate wall mount diagram 7 7 2 Installation with wall mount bracket diagram 8 8 HT16Ew Menu Operation Flowchart 9 8 1 Run mode description 16 8 2 disp Display par...

Page 3: ...he preferred protocol for low power devices and the transmission frequency is on the license free 2 4 GHz the transmission range is 90 meters typically without line of sight the data is pushed by the...

Page 4: ...l Strength Display type Segment LCD Display Character Size 7 segments with symbols character size 10 5mm MOC Enclosure Self extinguishing ASA UL 94 V0 White gray color 90 H x 130 W x 25 D mm Mounting...

Page 5: ...r s manual Page 5 of 26 5 HT16EwINTERNALBLOCKDIAGRAM HT16Ew UNIT Fig 1 Block Diagram of HT16Ew Micro Controller Real Time Clock RTC Digital humidity Temperature Sensor Data Flash LCD driver LCD displa...

Page 6: ...Ref No m94A OM 201 Issue No 01 Operator s manual Page 6 of 26 6 HT16EwAPPLICATIONDIAGRAM...

Page 7: ...Ref No m94A OM 201 Issue No 01 Operator s manual Page 7 of 26 7 HT16EwMOUNTING 7 1 Installation with Back plate wall mount diagram...

Page 8: ...Ref No m94A OM 201 Issue No 01 Operator s manual Page 8 of 26 7 2 Installation with wall mount bracket diagram...

Page 9: ...f 26 8 HT16EwMenuOperationFlowchart Main Menu Main menu is divided in main six categories 1 disp display parameters 2 P set parameter set 3 xbee Xbee Radio menu 4 excp exception parameters 5 cldr cale...

Page 10: ...Ref No m94A OM 201 Issue No 01 Operator s manual Page 10 of 26...

Page 11: ...Ref No m94A OM 201 Issue No 01 Operator s manual Page 11 of 26...

Page 12: ...Ref No m94A OM 201 Issue No 01 Operator s manual Page 12 of 26...

Page 13: ...Ref No m94A OM 201 Issue No 01 Operator s manual Page 13 of 26...

Page 14: ...Ref No m94A OM 201 Issue No 01 Operator s manual Page 14 of 26...

Page 15: ...Ref No m94A OM 201 Issue No 01 Operator s manual Page 15 of 26...

Page 16: ...eless communication established between HT16Ew CDU 85xx wireless symbol appear on top right corner of LCD if by chance communication link break then this symbol disappear In addition to that when HT16...

Page 17: ...to display quality of received signal strength of communication frame It s simply measure and displays received signal strength of last received packet Assume display digit in decibel dB RSSI 70 dB c...

Page 18: ...5xx Humidity High Alarm indication appear as symbol h ahead of humidity reading during run mode 5 DTMR Display Update Timer Mode DTMR is LCD Display update Timer mode It is user input menu user can si...

Page 19: ...ppear If user will press S n INC Key then it skip current mode If user will press S Y DCR Key usb d onscreen will appear which shows unit is ready to communicate with mCRM Tool After that connect HT16...

Page 20: ...Z is xbee PAN Personal Area Network ID It is user input menu user can simply enter 4 Digit PAN ID ranging from 0000 to 9999 with the help of 3 operating keys After setting PAnZ value device will lock...

Page 21: ...e Tbnd mode used to configure temperature band for exception protocol of transmitter unit Suppose Tbnd is set to 0 5then frame will only transmits when current temperature deviate 0 5 C from last meas...

Page 22: ...will press S Y DCR Key then it will reset FCtr to 0000 If user press DCR key in between it will shows the main menu screen 4 Lo9 oFF on Log Active Status Mode In case of communication break with CDU 8...

Page 23: ...e This mode shows the current month End user also change the month for unit On the press of INC key screen will appear which will show the month for current time To change the date user first press ME...

Page 24: ...key the first digit will blink Now using INC key user can change first digit Now if user press the DCR key cursor moves to second digit Now using INC key user can change second digit On the press of M...

Page 25: ...unit Step 1 Before applying new batteries check low battery status on display Low battery status indicated by single battery bar followed by min max status Step 2 First of all remove HT16Ew unit from...

Page 26: ...nicating on Wireless RF 2 4 GHz Zigbee Protocol with central Logger Unit Transmitter is battery operated low power unit which is sending data on Push time interval defined by user to Logger unit Time...

Reviews: