Strain Guage Indicator: 409-W
REF NO: m59A/om/501
Issue NO: 03
Page 34 of 34
Net Value = Gross Value- Tare Value
Example: If Input Range is 0-20mV, selected 30mV group. If Scaled Display Range is
0.00-40.00Kg., then I/P Zero will be 0.00 and I/P Span will be 40.00 with resolution
0.00.
If you feed input of 15mV then the Gross Value is as below:
Gross Value = 15.00* ((40.00-0.00)/ (20.00-0.00))
Gross Value = 15.00*2.00
Gross Value = 30.00
11.5 What is tare weight?
It is the weight of an empty object.
With an empty weighing device setting a zero point.
For the whole system is possible via the input Parameterised as tare under "Trigger
Input". The characteristic of the weighing cell is retained and only the origin is
Adjusted.
11.6 What is step value?
The step for the right most digit of display change is set with step value parameter
in menu mode.
Right most digit
of
Process
value
change
accordingly
Step: off
0
1
2
3
4
5
6
7
8
9
10
Step: 2
0
2
4
6
8
10
Step : 5
0
5
10
Step : 10
5
10