Strain Guage Indicator: 409-W
REF NO: m59A/om/501
Issue NO: 03
Page 25 of 34
• Sensitivity: 2mV/V
• Maximum load: 4000.0kg
Set calibration parameters as shown below:
SPAN = 4000.0 kg
MV-FS = 20.000 mV
Zero weight is empty scale weight.
Enter into
CALZ
mode with front panel keys,
Zero Calibration: Feed empty scale mV and Press
key
3. If it does not happen please check mechanical conditions and connection with
Indicator.
Note: calibrate unit with higher resolution for better result.
For Ex.
With Resolution 4 (9.9999 kg), Resolution 3 (99.999 kg), Resolution 2 (999.99 kg),
Resolution 1 (9999.9 kg), Resolution 0 (99999 kg).