6. Configuration Guidelines
6.1
Menu Parameter List
Display
Name
Description
Default
Value
RESO
(
RESO
)
Resolution
Set position of Decimal Point on Display.
0 :
0
1 :
.0
2 :
.00
3 :
.000
4 :
.0000
.
0
MAXMV*
(
MAXMV
)
Maximum Mv
range
Set mV range for resolution.
0:
15mV
1 :
30mV
2 :
75mV
30 mV
OFST
(
OFST
)
Offset
Set offset value for weight
-9999
to
9999
0. 0
KTARE
(
KTARE
)
Keypad Tare
option
Set value to enable or disable Key tare
0:
Disb
1:
En
eN
ZTARE
(
ZTARE
)
Zero Tare
set to clear Tare
0:
NO
1:
YES
NO
STEP
(
STEP
)
Step value
Set steps for display increment
0:
OFF
1:
2
2:
5
3:
10
OF F
Rtr
Retransmission
RET V
(
rETV
)
Retransmission
voltage
Retransmission Output Type
This output is according to PV input.
Zero & Span acts as Min & Max value of retransmission o/p
scale respectively.
0 :
0-10V
1 :
0-5V
2 :
1-5V
Voltage or Current is Jumper Selectable from the
Hardware.
0-10