Temperature Indicator: 409-4IN
REF NO: m47/om/201
Issue NO: 04
Page 28 of 34
Exceptional Response
CODE
MEANING
01
Function code Invalid. It must be 01, 03, 04,05 or 16.The function code received in the query is not
allowable action for the slave.
02
Illegal address value. The data address received in the query is not an allowable address for the salve.
03
Illegal data value. A value contained in the query data field is not an allowable value for the salve.
06
When Master device write some parameters to Slave device, If slave device busy then it will send 06
code to indicate slave device is busy.
Table 14.
9.
Technical Specifications
Display:
PV: Red LED 4-digit,
character size 4.00”.
LED for status indication (Alarm and Tx/Rx).
Operation keys: Menu, Enter, Increment, Decrement.
Input:
Input
Type
Range
Accuracy
TC
E
-200 to 1000 ºC
±0.1 % Of Full Span
±1Digit
(B,R,S,0-
6KΩ:
±0.15% Of Full Span + 2
Digit)
J
-200 to 1200 ºC
K
-200 to 1350 ºC
T
-200 to 400 ºC
B
450 to 1800 ºC
R
0 to 1750 ºC
S
0 to 1750 ºC
RTD
Pt 100
-200 to 850 ºC
DC
Current
4-20 mA
-1999 to 9999,
-1999 to 9999,
-1999 to 9999,
0-20 mA
0-5 V
1-5 V
0-2 V