User Manual
21
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
8.4.2.2 Configuration Mode Parameters Of 1008S With Mass Flow
33
Over Range
If Yes then integration total will be
running according to process value
corresponding to 20ma and if No then
integration total will stop after over limit is
reached.
-
Yes/No
34
Pulse time
For averaging of input pulse.
2
0 To 20
35
Frequency
Selection
For Frequency Lower than 300Hz select
frequency type low.
-
Low, High
36
Pulse width
To set width of input pulse in low
frequency mode.
2
0 To 50
Sr.
No.
Name
Description
No Of
Digit
Lo Limit
Hi Limit
1
Batch Mode
Type Of Mode, For Flow Control
-
Normal, Counter
2
Input Type
Type Of Input
-
0-5V, 1-5V, 0-20mA, 4-20mA ,
Pulse
3
Type Of
Instrument
Type Of Instrument, Which You Want To
Use.
-
Indicator/
Totaliser
4
Square Root
Mode Of Linearization
-
Yes/No
5
Digital Input
To Be Used To Reset Batch Total,
Integration total & to start batch and to
stop batch
Yes/No
6
Digital Filter
If Yes Than In Programming Mode It Will
Show Filter No.
-
Yes/No
7
Time Base
For Calculation/Display Of Flow Rate
-
Sec, Min, Hour, Day
8
Chnl1 ZR
Zero Value
5
00000
99999
9
Chnl1 FS
Full Scale Value
5
00000
99999
10
No Of
Batches
2
00
99
11
Batch Count
Counter Will Count How Much Batches
Has Been Taken
2
00
99
12
Relay Mode
Two Type Of Mode
-
Normal, Failsafe
13
Cut Off
3
000
100
14
DP
Decimal Point
-
0,0.1,0.01,0.001
15
SF1
Full Scale For Channel 1
4
0000
9999
16
SF2
Full Scale For Channel 2
6
00.00
99.99
17
SF3
Full Scale For Channel 3
6
000.0
999.9
18
Pcns
Pressure constant
5
000.00
999.99
19
tcns
Temperature constant
5
00000
99999
20
F2E
Channel #2 Linearization mode
-
Sqrt / No sqrt
21
ConE
Channel #3 Linearization mode
-
Sqrt / No sqrt
22
Zp
Zero setting for Channel-2
5
000.00
999.99
23
Zt
Zero setting for Channel-3
5
0000.0
9999.9
24
F1E
Channel #1 Linearization mode
-
Sqrt / No sqrt
25
Baud Rate
-
9600,19200
26
Serial No
2
01 TO 99
27
Pout
4
0001 TO 9999
28
Default
Display
-
Batch Total/Integration Total
(Bat Tot/Int Tot)
29
Pass
4
0000-9999
30
Conversion
Factor
4
00.00-99.99
31
Clear
Total
To Clear Integration /Batch Total And
Roll Count
4