Damping the AFC reaction
6.3.4
Ready / Start connections between UNITY–Air Server-xr and Shimadzu GC
The remote cable connections between UNITY–Air Server-xr and the GC are shown below. The UNITY–Air
Server-xr GC cable (SERUTD-5095) has a 25-pin D-type connector at the UNITY-xr end and free wires at the
GC end. Four wires need to be inserted into the connection block at the rear of the QP2010 as shown
below.
When connected to the Shimadzu GCMS system the UNITY–Air Server-xr ready (in) logic must be set to
Open as shown below. To set the logic select view>options from the drop down menus to open the options
dialogue box.