Loosen the screw at the bottom of the trap
pneumatics assembly.
Disconnect the 2 solenoid connections. Gently pull
the trap pneumatic assembly towards the front of
the instrument.
Once clear of the screw, pull the pneumatics up
and off the brass guiding block and rotate to the
right.
1. Place the trap alignment tool into the cold trap
box.
2. The tool should pass smoothly up to the o-ring
in the heated valve before further gentle pressure
moves the tool another 1-2 mm.
If the tool does not move smoothly past the
point where it leaves the cold trap box there
may be a misalignment problem.
3. Remove the trap alignment tool
NEVER SWITCH UNITY-xr ON WITH THE
TRAP ALIGNMENT TOOL INSTALLED
1-2 mm past
o-ring