5.
Place a second washer over the UNITY–Air Server-xr transfer line so that it can be secured against
the inside wall of the GC oven.
6.
Fit the clamp (
SERUTD-1125
) onto the transfer line, followed by the spacer (
SERUTD-1232
)
secure with the M6 nut supplied with the transfer.
Option 2 – Through Heated Injector Port
NOTE:
This option is only suitable if the inlet will not be required at any time.
1.
Remove components from the injector (e.g. septum, liner etc.) to enable the transfer line to pass
through unobstructed.
2.
As the entry hole will generally be larger than the diameter of the metal line sleeve, one or more of
the large washers supplied will be needed.
Oven wall
Oven wall
Spacer
Clamp
Washer
Washer
Heated Transfer Line
Nut
Oven wall
Oven wall