Series E Ergonomics Force Gauges
User’s Guide
9
the desired value has been reached, press
ENTER
to save the change and revert back to the sub-menu
item, or press
ESCAPE
to revert back to the sub-menu item without saving. Press
ESCAPE
to revert one
step back in the menu hierarchy until back into normal operating mode.
Refer to the following sections for details about setting up particular functions and parameters.
5 SET POINTS
5.1 General Information
Set points are useful for pass / fail testing. Two limits, high and low, may be specified. The gauge
compares the primary reading to these limits, providing “under”, “in range”, and “over” indication.
5.2 Configuration
To configure set points, select
Set Points
from the menu. The screen appears as follows:
One, two, or none of the set points may be enabled. To toggle between the tension and compression
directions, press the
DIRECTION
key.
If two set points have been enabled, they are displayed in the upper left corner of the display. If only one
set point has been enabled, the word “OFF” appears in place of the value. If no set points have been
enabled, the upper left corner of the display will be blank.
When set points are enabled, the following indicators are shown to the left of the primary reading:
– the displayed value is greater than the upper
force limit (NO GO HIGH)
– the displayed value is between the limits (GO)
– the displayed value is less than the lower force
limit (NO GO LOW)
Note:
Set point indicators and outputs reference the displayed reading, not necessarily the current live load.
SET POINTS
Upper Disabled
* Upper Enabled
50.00
lbF
Lower Disabled
* Lower Enabled
35.00 lbF