*Reduced capacity:
24 l/min, 5 °F (3 °C) [temp. difference] at 50
psi (350 kPa/3.5 bar). This reduced capacity
applies if one or more of the limitations above
apply.
Expected flow and dew point suppression
Note: At a continuous flow of >30 l/min (normal for an adult patient is 8–10 l/min) a surrounding
temperature of approximately 104 °F (40 °C), and a relative humidity of >90% RH, the pressure
dew point could be less than 5 °F (3 °C).
Environmental conditions – Transport and storage
-40 °F to +158° F (-40 °C to +70 °C)
Ambient temperature
10% to 95%
Ambient relative humidity
500 to 1 200 hPa (375 to 900 mmHg)
Ambient atmospheric pressure
24
COMPRESSOR MINI, User´s Manual
English version
28.0
9. Technical data