Troubleshooting
Reason and action
Malfunction
•
Check that the mains cable is correctly plugged into
the mains in/outlet.
•
If it still does not start, fuses may be blown. Only
authorized personnel may check and replace fuses.
The compressor does not start; the main LED is not lit.
•
Too high air consumption. The compressor cannot
supply the connected equipment. Check the tubing
for leakage.
•
If the alarm continues contact authorized service
personnel.
Pressure alarm is activated.
•
Ambient air temperature is too high. Run at a cooler
temperature.
•
The air vents at the back or on the underside of the
compressor are covered or blocked. Remove
obstacles.
•
The dust filters are clogged. Clean or replace.
•
Cooling fan or dryer malfunction. Contact authorized
service personnel.
Temperature alarms activated.
The cooler/dryer system is not performing correctly.
Contact authorized service personnel.
Abnormally large quantities of water flush out from the
drain valve when the valve pin is depressed.
COMPRESSOR MINI, User´s Manual
English version
28.0
21
8. Troubleshooting
Summary of Contents for COMPRESSOR MINI
Page 1: ...User s Manual VENTILATION ANESTHESIA COMPRESSOR MINI...
Page 2: ......
Page 4: ......
Page 8: ...6 COMPRESSOR MINI User s Manual English version 28 0 1 Before use...
Page 14: ...12 COMPRESSOR MINI User s Manual English version 28 0 3 Preparation Connection...
Page 18: ...16 COMPRESSOR MINI User s Manual English version 28 0 5 Preventive maintenance Storage...
Page 20: ...18 COMPRESSOR MINI User s Manual English version 28 0 6 Cleaning...
Page 22: ...20 COMPRESSOR MINI User s Manual English version 28 0 7 Alarms...
Page 24: ...22 COMPRESSOR MINI User s Manual English version 28 0 8 Troubleshooting...
Page 27: ......