35
1133.02BX
GA113302GB11
VII. Technical data
1. Operating table adjustments
Length without head rest:
1900 mm
(74.8 inches)
Length with head rest:
2196 mm
(86.5 inches)
Width:
530 mm
(20.9 inches)
Width across side rails:
max. 580 mm
(22.8 inches)
Height (without pads):
594 - 1056 mm
(23.4 – 41.6 inches)
Weight:
approx. 328 kg
(723 lb.)
Trendelenburg position:
max. 30°
Reverse Trendelenburg position:
max. 30°
Lateral tilt left / right:
max. 20°
Lower back plate up:
max. 80°
Lower back plate down:
max. 40°
Upper back plate up:
90°
(manual adjustment)
Upper back plate down :
45°
(manual adjustment)
Leg plate up:
10°
Leg plate down:
90°
Longitudinal shift:
230 mm
(9 inches)
Flex position:
max. 40°
Reverse Trend. position:
20°
Back plate down
20°
Beach chair:
Trendelenburg position:
25°
Back plate up
80°
Leg plate down
70°
Max. patient weight:
450 kg (1000 lb)
2. Electrical data
Nominal voltage:
AC100/110-115/127
200/220/230-240V
Frequency:
50/60 Hz
Power consumption:
max. 400 VA
Class:
II (Double Insulated)
Applied part
Type B
The enclosure leakage current meets the requirements
of the patient leakage current for CF conditions ac-
cording to IEC 60601-1
IP X4* IPS ** SELV*** DC 24V
Duty cycle:
int. 10 min. ON / 20 min. OFF
For patients weights from 225 kg up to 450 kg
(500-1000 lb): 2 min. ON / 120 min. OFF
Explosion protection:
AP for operation on battery power
When connected to the mains, the operating table is
not approved for areas subject to explosion hazards
(AP-M)
Rechargeable batteries:
Type A512C / 15SR
2 rechargeable batteries 12 V / 15 Ah
*
IPX 4 =
protection against ingress of liquids
**
IPS
=
internal power source
***
SELV =
safety extra low voltage
VII. Technical data