22
1133.02BX
GA113302GB11
Î
Î
Î
Î
Î
Note
We recommend to recharge the batteries overnight,
but at least every 2 days, even if “low charge level”
of the rechargeable batteries has not been indicated
yet.
Regular recharging of the batteries will not only en-
sure sufficient battery power available for the opera-
tion programme, it will also serve to prolong the use-
ful life of the rechargeable batteries.
3. Operation on mains power
As soon as the operating table is connected to the mains
via the mains cable, the batteries are recharged auto-
matically.
There is no explosion protection while recharg-
ing the batteries or the operating table is on mains
power.
The separation of the operating table from the
mains is made by the mains plug.
Connect the operating table to the mains via the mains
cable.
When connecting and removing the mains ca-
ble, observe the correct order of the procedure:
•
When connecting the charging cable, first of
all connect the operating table and then plug
the cable into the socket.
•
When disconnecting the charging cable, first
of all pull the plug out of the socket and then
remove the cable from the operating table.
Î
Î
Î
Î
Î
Note
Damaged cables must be replaced immediately.
When replacing the fuses please note:
1.
Unplug the operating table
2.
Unscrew the two fuse holders located on the
right and left next to the socket for the main
plug with a coin or a large screw driver
3.
Insert the appropriate fuse according to the
opposite drawing
4.
Reinsert the fuse holders
4. HF surgery / Use of defibrillators
In the field of HF surgery and when using defibrillators,
be absolutely sure to observe the following safety in-
structions:
Danger of burns!
When using HF surgical equipment, defibrillators
and defibrillation monitors, make sure to avoid
that the patient comes into contact with the metal
parts of the operating table, the table tops or the
accessories. Also make sure to avoid that the
patient is lying on moist underlays or sheets on
the conductive pads.
Be absolutely sure to comply with the manufac-
turer’s operating instructions.
Socket for mains cable
Mains cable
Potential equalization
IV. Use of the operating table