10
1133.02BX
GA113302GB11
z)
0-position
Resets the lateral tilt, Trendelenburg/reverse Tren-
delenburg, back plate and leg plate adjustments to
the horizontal position. If the function key z is
pressed, the operating table top will first reset the
adjustments in the following order – lateral tilt, Tren-
delenburg/reverse Trendelenburg, back plate. An
acoustic signal will indicate that these adjustments
have been reset. Release the function key z and
press once again to reset the back plates to the
horizontal position; if the leg plates have been ad-
justed individually, they will be positioned parallel
to one another.
Î
Î
Î
Î
Î
Notes:
Manual adjustment functions such as those for the
upper back plate, head rest and leg plate abduction
will not be affected by the IR remote control (and as
a consequence neither by the automatic 0-position
function).
Î
Î
Î
Î
Î
The transmission power of the IR hand control is
sufficient to be able to adjust the operating table
even from a greater distance. If the operating table
does not respond to the command even though the
function key is completely depressed, either reaim
the hand control or move to a slightly different trans-
mission position. In such cases the receiver at the
column is probably being blocked by a person’s
body.
2.2. Override system = emergency operation
2.2.1 Override control panel
In case of a malfunction or a defective hand control, you
can control the operating table functions via the control
panel 13 on the upper column casing = emergency op-
eration.
Î
Î
Î
Î
Î
Note:
Always press the ON key together with the desired
function key.
Î
Î
Î
Î
Î
Note:
If the operating table batteries are discharged, the
override system only works if the operating table is
connected to the mains.
Î
Î
Î
Î
Î
Note:
When using the override control panel, the
anticollision feature is not activated. Avoid extreme
adjustments – the combination of maximum Tren-
delenburg/reverse Trendelenburg position and maxi-
mum lateral tilt.
13
III. General description