10 EQUIPMENT DECOMISSIONING
Before dismantling and scrapping the equipment, disconnect the power
supply by unplugging the machine.
The equipment decommissioning process must be performed by qualified
personnel with appropriate PPE (personal protective equipment) as laid down
by current regulations.
The equipment contains electronic components, electromechanical components,
plastic and metal parts. Each part must be disassembled and disposed of separately
in accordance with the laws in force on the disposal of industrial solid waste.
11 PROBLEMS AND SOLUTIONS
31
PROBLEM
CAUSE
SOLUTION
The machine does not turn on
Main switch at the back of the machine off
Turn on main switch
Key switch under the touch screen off
Turn on key switch
Power cable not plugged in properly
Insert plug fully
Circuit breaker activated
Contact Technical Assistance
Blown fuse
Contact Technical Assistance
Touch-screen control panel displays
indefinite images
Temporary unknown cause
Reboot the machine using the key switch.
If the problem persists, contact Technical
Assistance
No flashing light on LED when the
Maxenergy handpiece should be active
Connector not inserted properly or
damaged
Insert connector (see par. 6.4.2 Assembly)
or replace if damaged (see par. Spare
Parts)
Damaged software
Contact Technical Assistance
No flashing light on LED when the Vixo
handpiece should be active
Damaged software
Insert connector (see par. 6.4.2 Assembly)
or replace if damaged (see par. Spare
Parts)
Damaged software
Contact Technical Assistance
Failure of the vacuum system or weak
vacuum effect on one of the handpieces
Suction hose is not inserted properly
Insert suction hose (see par. 6.4.2
Assembly)
Clogged vacuum filter
Clean filter (see par. 8.1.3 Vacuum filters
cleaning and cartridge replacement)
Foreign body stuck in the handpiece
If it is not visible, contact Technical
Assistance
Damaged software
Contact Technical Assistance
Failure of the vacuum system on all the
handpieces
Damaged software
Contact Technical Assistance
Suction hoses are not inserted
Insert suction hoses (see par. 6.4.2
Assembly)
Vacuum pump is broken
Contact Technical Assistance
When the machine is turned on, the LEDs
on the plexiglas do not switch on
Connector is not inserted properly
Insert connector (see par. 6.4.2 Assembly)