7.4
Repeat a treatment cycle
It is possible to repeat a treatment
cycle previously stored (SAVE USB)
by inserting the USB stick containing
all the required data and then select-
ing the USB item, under menu (F).
Select USB and then FORWARD.
All data should be
carefully stored in
order to make their
recovery easy and
safe.
Clean and disinfect the handpieces before starting a new treatment (see paragraph
on cleaning and disinfection of the equipment)
7.5 Interruption of the ongoing treatment
It is possible to stop the ongoing treatment (pause) at any time by quickly pressing the
START/STOP button on the handpiece or clicking the PAUSE button on the treatment menu
displayed on the touchscreen. To resume the treatment, press both the START button on the
touch screen and the START/STOP button on the handpiece (for 1 second).
By pressing the STOP button on the touch screen while a treatment is in progress, the timer is
reset to zero.
7.6 Equipment emergency stop
All equipment functions can be immediately stopped at any time by pressing the emergency stop
button which is located under the touch-screen control panel of the machine (see 4.1 Equipment).
7.7 Equipment recovery in case of emergency
Once the problem causing the emergency has been solved, reset the emergency stop button by
rotating it clockwise. MR991 will switch on automatically, restarting from the initial self-test page.
If a treatment has been interrupted, set its remaining duration time.
7.8 Recovery from power failure
After a power failure, MR991 will switch on automatically, restarting from the initial self-test page.
If a treatment has been interrupted, set its remaining duration time.
27