Mai ntenance
69
To remove any blockages in the paddle stem and clean the inner diameter of the glass syr-
inge, follow the instruction video and/or the software guided workflow integrated in the
MicroCal PEAQ-ITC Control Software.
The software guided workflow is also shown in the following table.
Step
Action
1
Carefully insert the supplied Cleaning Wire into the end of the paddle and pass it up
through the stem into the syringe glass.
2
Scrub the inside of the syringe with detergent using the supplied Syringe Glass Cleaning
Brush.
3
Rinse thoroughly with water to remove as much detergent as possible.
4
Dry the syringe to prevent sample dilution.
Install the syringe glass
To install the syringe glass, follow the instruction video and/or the software guided work-
flow integrated in the MicroCal PEAQ-ITC Control Software.
The software guided workflow is also shown in the following table.
Step
Action
1
Grasp the pipette's rotating assembly.
2
Gently insert the syringe glass with the syringe's fill port facing the same direction as the
hole in the pipette's rotating assembly.