H ealth and safety
13
!
Warning! Liquid solutions
. Use caution when using solutions near the instru-
ment. If any liquid is spilled on or around the instrument, unplug the instru-
ment immediately and wipe up the spilled liquid. If there is any possibility that
liquid may have leaked into the instrument case, contact Malvern Instruments
immediately. Do not plug the instrument into any electrical outlet until the
problem is resolved.
!
Warning! Strong acids
. The MicroCal PEAQ-ITC cells are constructed out of
Hastelloy™. Strong acids must be avoided.
!
Notice! Electromagnetic and radio frequency interference (EMI/RFI).
Lab-
oratory tests have shown that significant electromagnetic waves may affect the
performance of MicroCal PEAQ-ITC.
Strong electrical or magnetic fields may be produced by an NMR, microwave
oven, large motors or refrigeration units. In some cases, these waves can cause
unintended data results.
Please refrain from all use of such equipment.
Using flammable liquids
!
Warning! Fire Hazard
. Before starting the system, make sure that there is no
leakage.
!
Warning!
A fume hood or similar ventilation system shall be installed when
flammable or noxious substances are used.
Personal protection
!
Warning!
Always use appropriate Personal Protective Equipment (PPE) during
operation and maintenance of this product.