48
|
Fig. 2 (dimensions of a split side rail)
Check whether the required distances are also complied with
under load. Check dimensions A and D using a test tool in line
with BS EN 60601-2-52. The test force for dimension A is 250 N.
Dimensions
Requirement
[mm]
A
The largest dimension in at least one
direction between components of the
side rail / handrail in all commonly used
positions.
A ≤ 120
B
Thickness of the normally used mattress
without compression as stated by the
manufacturer.
As stated
by the
manufacturer
C
Distance from the upper edge of the side
rail to the upper edge of the mattress
without compression with the sleeping
surface in a level position.
C ≥ 220
D
Distance between head / foot board or ac-
cessories to the side rail / handrail is parallel
to the sleeping surface in a level position.
This also applies to beds with extended
foot section.
D ≤ 60 or
D ≥ 318
E
Distance between split side rails with
sleeping surface in a level position.
E ≤ 60 or
E ≥ 318
F
The largest dimension in at least one direc-
tion of every opening below the side rail.
If D ≥ 250,
F ≤ 60
If D ≤ 60,
F ≤ 120
G
Length of the side rail(s)
G ≥ 2 / 3 H
H
Inner distance between head and foot
board without any extension of these
parts
No
requirements
3. Functionality check of brakes
Depending on the variant, the functionality of the brakes must be
checked in every position. With electrical brake systems, check
whether the brakes fully retract and extend.
4. Functionality check of lifting motors
Move all lifting motors to their end position and back again. When
doing so, please observe the following points:
§
Any unusual noises generated
§
Synchronism of the lifting motors
§
Smooth operation of the lifting motors
§
Correct path of the lifting motors
§
Automatic switch-off in the end position
The travel path of the lifting motors may vary depending on the
model variant. If in doubt, please contact our customer service.
Summary of Contents for IMPULSE KL
Page 1: ...CLINIC DESIGN IMPULSE KL INSTRUCTION MANUAL HOSPITAL BED...
Page 2: ...2...
Page 13: ...13...
Page 31: ...31...
Page 33: ...33 GS side rail dimensions When using the ABS metal lattice sleeping surface 378 114 85...
Page 36: ...36 VGS side rail dimensions When using the ABS metal lattice sleeping surface 368 115 85...
Page 41: ...41...
Page 62: ...62 Notes...
Page 63: ...63 Notes...