Front Loading Utensil Washer Disinfector (WDF)
Operation, Maintenance and Installation Manual
Issue 1
Page
17
20/01/2022
Fault 34
Level Sensor Fault
(Level sensor fault detected)
Cause
Split tube to pressure sensor
Water build up in pressure sensor tube
Pressure sensor failed
Fault 35
Control Temp Probe Fail
(Probe fault detected)
Cause
Broken or shorted out connection
Probe failed internally
Fault 36
Validation Temp Probe Fail
(Probe fault detected)
Cause
Broken or shorted out connection
Probe failed internally
Fault 37
Temp. Probe Differential
(Control & validation temperature probe measurement discrepancy)
Cause
Probe not installed correctly
Probe fowled or covered by foreign object
Intermittent probe fault
Probe Failure
Fault 40
Master Reset
(A master reset of the device has been performed)
Cause
Service intervention