Operating Instructions
35
8.3 Display Contrast
To change the contrast of the display, select
Display
from the set up
menu.
Use the up
and down
arrows to adjust the contrast. When done
adjusting the contrast press the right
arrow button to
Save
and exit.
8.4 Instrument Details
To view the instrument details, select
Instrument Details
from the set
up menu. Instrument details will show the serial number of the
instrument, connected probe, firmware (embedded operating
software) version and a summary of other instrument settings.
8.5 DP Early Stop
The user can select how the instrument will operate when performing
a DPOAE test. With the
DP Early Stop feature off
, testing will be
completed at all frequencies included in the selected protocol. The test
is complete when all frequencies have been measured. If the
DP Early
Stop
feature is
on,
the measurement will stop as soon as the
instrument can determine that the criteria for a PASS or REFER result
has been met. In this case some frequencies may not be measured,
but testing will be completed faster.
To change this setting, select
DP Early Stop
from the
Set Up
menu.
Using the up
and down
arrow buttons on the control panel select
off
or
on
and then press the right
arrow button to
select
and exit.
Summary of Contents for Ero Scan Pro
Page 74: ...1162 0802 Rev E 05 12...