41
BA315001-en
Error
code
Error/Error message
Diagnosis/Corrective action
12
Check filter;
pump defective
Are F1 filters mounted correctly? Is the cover of the water
extractor open?
Master switch off, wait 10 s, master switch on.
No improvement?
Notify Service.
13
Filter blocked!
Change both F1 filters.
Master switch off, wait 10 s, master switch on.
No improvement?
Notify Service.
14
Heater/temperature
sensor defective
Master switch off, wait 10 s, master switch on.
No improvement?
Notify Service.
15
Water extractor/
temperature sensor de-
fective
Master switch off, wait 10 s, master switch on.
No improvement?
Notify Service.
16
Ambient temperature/
sensor defective
Master switch off, wait 10 s, master switch on.
No improvement?
Notify Service.
17
WiFi password is incor-
rect
Check and adjust the unit’s WiFi settings as described in
the section entitled ‘PC–MET connection options / WiFi
connection’.
No improvement?
Check WiFi router, settings altered?
Notify Service.
18
Error occurs during
leaktightness test
Probe seal tubing correctly pushed on to probe tip?
Sample line inserted into unit?
Are the two F1 filters mounted correctly?
Are the transmitter and receiver of the opacimeter
mounted correctly?
Is the cover of the water extractor screwed on correctly?
Master switch off, wait 10 s, master switch on.
No improvement?
Notify Service.
19
Limit value exceeded in
HC residue test
Ensure that the sampling probe is placed such that no
hydrocarbons (such as those contained in the exhaust
gas or brake cleaner) are induced and re-start the test.
Rinse out the unit by running it in measurement mode
for 2-3 minutes. Ensure that the probe does not take in
any hydrocarbons.
Change both F1 filters.