Error
code
Error/Error message
Diagnosis/Corrective action
04
Water extractor faulty,
power consumption too
high
You can acknowledge errors by clicking the ‘Confirm’
button. Measurement is possible, but be aware that there
is a high risk that the unit will fail! – Better:
Master switch off, wait 10 s, master switch on.
No improvement
Notify Service.
05
Heater faulty
Master switch off, wait 10 s, master switch on.
No improvement
Notify Service.
06
O
2
sensor defective/old Change O
2
sensor as described in op. manual
07
LED too dim
Trans receiver ins screwed in?
Install
Glasses soiled?
Clean
Master switch off, wait 10 s, master switch on.
No improvement?
Notify Service.
08
Opacimeter sensor de-
fective
Trans receiver ins screwed in?
Install
Filter glass holder inserted correctly?
Check
Servicing hatches closed during start-up of unit?
Close hatches
Glasses soiled?
Clean
Master switch off, wait 10 s, master switch on.
No improvement?
Notify Service.
09
Reference sensor or
LED defective
Trans receiver ins screwed in?
Install
Glasses soiled?
Clean
Master switch off, wait 10 s, master switch on.
No improvement?
Notify Service.
10
Adjustment deadline
expired
Adjust infrared test bench as per op. manual
11
Gas zero-balance ad-
justment failed
Master switch off, wait 10 s, master switch on.
No improvement?
Check that filter F2 is installed/free
from blockage
Check error codes
No improvement?
Notify Service.