16
BA010101-en
6
Maintenance
DANGER
Risk of death or severe personal injury by electric shock
Before doing any maintenance work, turn off the main switch and protect it
against tampering.
6.1
Maintenance Schedule
Interval
Maintenance Items
Procedure
3 months
Hydraulic system
Check fluid level, refill if necessary.
Check hydraulic system for leakages.
Check power unit for unusual noise dur-
ing operation, check fastening screws
for tight fit.
6 months
Hydraulic fluid
Check fluid for contamination and aging,
replace if necessary.
12 months
General inspection
Check all components for damage.
6 years
Pressure hoses
Replace pressure hoses
6.2
Annual Inspection
•
The maintenance interval prescribed by the manufacturer is 12 (twelve) months.
This maintenance interval refers to normal workshop usage. If the equipment is
used more frequently or under severe operating conditions (e.g. outdoors), the
interval must be reduced accordingly.
•
Maintenance work shall be done only by authorized and trained service techni-
cians provided by the manufacturer, licensed dealers or service partners.
•
In case of non-compliance the manufacturer's warranty becomes void.